Overview Specifications Documentation & Tools ;
Overview

Overview

Product Features
  • 2x12 round cells (16 mm) for 2 simultaneous digestion programs
  • 8 pre-programmed settings including fast digestion for COD
  • Easy, quick operation for standard parameters
  • AQA/IQC with external temperature sensor

The WTW Thermo Reactor CR 4200 is a dry temperature control device for use in the laboratory. It facilitates and safeguards the digestion with reaction cells.

Two different programs for 12 cuvettes each can be run at the same time. You can use an external temperature sensor instead of a vial with data output to the PC/printer for analytical quality assurance (AQA).

8 standard digestion programs are included for the standard parameters COD, quick COD (148°C, 20 min.), total nitrogen, total phosphorus, TOC and heavy metals of galvanizing with automatic heating and switch-off.

8 user-defined digestions at temperatures between 25-170°C can be programmed for special digestions allowing also high temperature digestions.

Programs for standard parameters:

  • 1: 148 °C for 120 minutes
  • 2: 120 °C for 30 minutes
  • 3: 120 °C for 60 minutes
  • 4: 120 °C for 120 minutes
  • 5: 100 °C for 60 minutes
  • 6: 148 °C for 20 minutes
  • 7: 150 °C for 120 minutes
  • 8: 100 °C for 30 minutes

SKU: 1P23-

Specifications

Specifications

Technical Specifications
Thermoreactor CR 4200
Reactor Types
Dry temperature control device with safety cover
Cell Shafts
2 x 12 cell shafts for reaction cells, diameter: 16 ± 0.2 mm
Reaction Time Setting
program settings or user-defined programs up to 180 min
Temperature Setting
stored program, user-defined settings from room temperature up to 170°C
Controlling Accuracy
± 1 °C ± 1 digit
Temperature Stability
± 0.5 K
Overtemperature Protection
190 °C ± 5 °C
Dimensions
D x W x H: 312 x 255 x 185 mm
Weight
4 kg
Output
unidirectional RS232 interface with 2 banana sockets for external temperature sensor (article-# 250 100), printer, PC
Power Supply
115/230 VAC 50/60 Hz switchable via fuse setting allowed voltage tolerance: ±15 % maximum power consumption: 560 W, Fuse 2x6.3 AT
Protection Class
IP 20 according to EN 60529
Ambient Temperature
Operation +5 °C to +40 °C
Test Certificates
cETLus, CE
Climactic Class
Thermoreactor CR 4200
Additional 1
2 according to VDI/VDE 3540
Additional 2
Relative humidity: Yearly mean: < 75 %
Additional 3
30 days /year: 95 % Other days: 85 % Light dew: yes
Additional 4
Site altitude: max. 5000 m above sea level
EMC
Thermoreactor CR 4200
Additional 1
EN 61326-1
Additional 2
FCC 47 CFR Part 15
Safety Standards
Thermoreactor CR 4200
Additional 1
EN 61010-1
Additional 2
EN 61010-2-010
Additional 3
UL 61010-1
Additional 4
CAN/CSA C22.2#61010-1
Additional 5
CAN/CSA C22.2#61010-2-010
Documentation & Tools