Overview Specifications Documentation & Tools ;
Overview

Overview

The WTW Thermoreactor CR 3200 has programs for standard parameters COD including fast digestion (148°C, 20 min.), total nitrogen, total phosphorus, TOC and heavy metals of galvanizing with automatic heating and switch-off quick. User defined programs can be easily entered and stored from room temperature respectively 25-170°C to perform special digestions and thus allow high temperature digestion.

Product Features
  • Digestions of 24 round cells with pre-programmed and user-defined programs
  • 8 stored programs for easy and fast operation including COD fast digestion
  • 8 user-defined programs from room temperature/25-170°C with high temperature digestion
  • AQA/IQC with optional external temperature sensor for documentation

The WTW Thermoreactor CR 3200 program can be run for all vials at the same time; to perform two parallel digestion programs, the CR 4200 is the suitable choice. For the analytical quality assurance (AQA), the external temperature sensor TFK CR can be connected to the PC/printer instead of a round cell.

Programs for standard parameters

  • 1: 148 °C for 120 minutes
  • 2: 120 °C for 30 minutes
  • 3: 120 °C for 60 minutes
  • 4: 120 °C for 120 minutes
  • 5: 100 °C for 60 minutes
  • 6: 148 °C for 20 minutes
  • 7: 150 °C for 120 minutes
  • 8: 100 °C for 30 minutes
Specifications

Specifications

Specifications
Thermoreactor CR 3200
Reactor type
Dry temperature control device with safety cover
Cell shafts
2 x 12 cell shafts for reaction cells, diameter: 16 ± 0.2 mm
Reaction time setting
program settings or user-defined programs up to 180 min
Temperature setting
stored program, user-defined settings from room temperature up to 170°C
Controlling accuracy
± 1 °C ± 1 digit
Temperature stability
± 0.5 K
Overtemperature protection
190 °C ± 5 °C
Dimensions
D x W x H: 312 x 255 x 185 mm
Weight
4 kg
Output
unidirectional RS232 interface with 2 banana sockets for external temperature sensor (article-# 250 100), printer, PC
Power supply
115/230 VAC 50/60 Hz switchable via fuse setting; allowed voltage tolerance: ±15 % maximum power consumption: 560 W, Fuse 2x6.3 AT
Protection Class
IP 20 according to EN 60529
Ambient temperature
Operation +5 °C to +40 °C
Climatic class
2 according to VDI/VDE 3540; Relative humidity: Yearly mean: < 75 %; 30 days /year: 95 % Other days: 85 % Light dew: yes; Site altitude: max. 5000 m above sea level
EMC
EN 61326-1; FCC 47 CFR Part 15
Test certificates
cETLus, CE
Safety standards
EN 61010-1; EN 61010-2-010; UL 61010-1; CAN/CSA C22.2#61010-1; CAN/CSA C22.2#61010-2-010